BioSAFE Engineering - A Clean Energy Company

150 lb/hr

150 lb/hr

150 lb per hour medical waste treatment unit

Throughput Capacity 150 lb (68 kg) per hour
Operating Temperature Waste in auger chamber: 205-212°F (96-100°C)
Steam in auger chamber jacket: 220-240°F (104-115°C)
Construction Material Stainless steel (optional painted steel)
Safety
  • Safety switches on all access panels
  • Siemens PLC monitored alarm matrix
  • Spore testing made easy
Electrical Enclosures Shredder: NEMA 4 painted mild steel.
Main: NEMA 4 with Siemens PLC and touchscreen
HEPA Filter Min. operational efficiency of 99.97@0.3 microns
Machine Footprint Aprx. 13.58 ft long by 5.5ft wide by 8.9 ft tall
Time in Auger Chamber ≥ 60 Minutes
Recommended Utilities Sewer: 1″ NPT Route to floor drain 3-4″ sink drain recommended
Ventilation: 4″ to outside above roof (2 vents) schedule 80 cpvc.
Steam: 60 lb/hr regulated to 14 psig
Electric: Shredder: 480 VAC, 3 phase plus ground, 60 amp, 60 Hz, fused disconnect main: 480 VAC, 3 phase plus ground, 30 amp, 60 Hz
Communications: Ethernet cat5 preferred or standard 4 wire analog with call access.
(All utilities within 10 feet of unit)
Options
  • Stainless steel or painted mild steel construction
  • Service contract
  • Additional training services